Autonomous Endovascular Navigation Using 2D Image Feedback
Autonomous Endovascular Navigation Using 2D Image Feedback
arXiv Robotics — research abstracts1Sheng Cao, Shuxiang Guo, Yonggan Yan, Jian Guo, Chunying Li, Bin Wang, Mingchao Ding, Yongxin Zhang, Yongwei Zhang, Pengfei Yang, Jianmin Liu
A perception and control framework uses 2D image feedback for autonomous endovascular navigation. Evaluated on a dual-device robotic platform, it achieved 0.95 guidewire progress in 65 runs, with 3.1% failing due to autonomous stop events.
arXiv:2610.21481v1 Announce Type: new
Abstract: In endovascular interventions, physicians rely on planar DSA and X-ray fluoroscopy to interpret vessel anatomy and device motion, yet converting these images into feedback for autonomous navigation remains difficult. A perception and control method is formulated for two-dimensional image feedback. Vessel masks are converted into a labeled centerline graph, and device-tip estimates are projected onto a planned path to obtain progress, deviation, and direction. A rule-based controller uses these quantities for delivery, branch selection, probing, rollback, and recovery. The method was evaluated on a physical dual-device robotic platform in 65 closed-loop runs using simulator-rendered planar images across three paths. All runs reached at least 0.95 maximum observed guidewire progress; two runs (3.1%) met the prespecified failure rule of three or more autonomous stop events. The results establish route-level performance on the tested platform; clinical image sequences and broader anatomy remain to be evaluated.