OUCH: Objective, Uniform, Controlled Haptic Stimulation for Automated Neurological Assessment
OUCH: Objective, Uniform, Controlled Haptic Stimulation for Automated Neurological Assessment
This research introduces OUCH, a wrist-worn device using pneumatic actuation to deliver consistent noxious mechanical stimuli for assessing neurological function.
The study presents a wrist-worn pneumatic device called OUCH that delivers controlled noxious haptic stimuli for neurological assessment, demonstrating higher contact pressure concentration and lower required force compared to a commercial handheld algometer in preliminary tests.
Source: arXiv Robotics — research abstracts · Read original article ↗
Loading article text…
What the source reports
Publisher-reported claims, with original evidence. These results have not been independently verified by RoboSignal.
- paper: not_reported
Source excerpts and review record
Automatically extracted; no manual editorial approval recorded.
arXiv:2611.05741v1 Announce Type: new Abstract: Motor responses evoked by noxious stimulation are used to assess neurological function in hospitalized patients with disorders of consciousness. In standard bedside assessments, these stimuli are applied manually, resulting in substantial variation in force, duration, frequency, and location. Here, we present OUCH (Objective, Uniform, Controlled Haptic stimulation), a w
Open source S2
rist-worn device that uses pneumatic actuation to deliver noxious mechanical stimuli. OUCH combines stacked pneumatic pouch actuators, a rigid indenter, and a compliant spacer designed to buckle at a set load, so that the indenter contacts the skin only during actuation. We characterize the device using blocked force tests, clinician evaluation, finite element analysis, and a deformable pressure testbed, and benchmar
Open source S3
k it against a commercially available handheld pressure algometer. At 200 kPa pouch inflation pressure, our indenter concentrated its load into an estimated contact pressure of 18,500 kPa, over 25 times that of the algometer tip, while requiring less total force. On the deformable testbed, it localized deformation near the surface, while deformation at depth remained comparable to the algometer. As a result, in preli
Open source S4
minary perceptual testing, our indenter elicited discomfort at approximately 89% lower applied force than the algometer, demonstrating the feasibility of evoking discomfort with a compact, wearable soft actuator. These results show that combining pneumatic actuation with designed contact geometry enables controlled, repeatable noxious stimulation at the wrist, providing a foundation for objective neurological assessm
Open source S5
ent.
Open source S6
OUCH: Objective, Uniform, Controlled Haptic Stimulation for Automated Neurological Assessment
Open source S1
Source:arXiv Robotics — research abstracts · arxiv.org